Increase shelf-life and prevent spoilage
High resolution ethylene measurement from 0-10ppm.
Controlled Atmosphere Facilities
Monitor ethylene mitigation and maintain the proper balance of gasses in controlled atmosphere environments.Ripening Rooms
Precision citrus degreening and ripening control on various commodities.Modified Atmosphere Packaging
QA assessment in packing lines, laboratories, import/export surveys, distribution centers and retail centers.Optimize your headspace and MAP analysis workflows.
Gather precise CO2 and O2 data in seconds with the Check It! Gas Analyzer.
Backlit LCD screen
Never struggle to see your readings, even in dimly-lit storage environments.
Lightweight & heavy duty
Individually machined in the USA, the Store It! Gas Analyzer’s highly durable, aluminum construction gives you peace of mind without sacrificing weight.
Automated data capture
Never lose a data point with up-to-the-second autosave.
Internal data logging
Store and review thousands of records on the Check It! Gas Analyzer’s removable 16gb SD card.
Request a live demonstration
Vital QA data collection in the palm of your hand
All-day battery life
With an 8.5 hour battery life and rechargeable 3400 mAh replacement batteries included, you can work confidently without interruption.
Data transfer
Rapid data transfer through SD card and wired connection.
Rapid measurement
With CO2 and O2 measurements in under 10 seconds, you won’t be stuck waiting for your data.
User-selected sampling modes
Choose from headspace or continuous sampling depending on your unique needs.
5 data points recorded
Each sampling event records CO2 concentration, O2 concentration, date & time, RH, and GPS location.
Wide temperature and humidity range
Gather reliable data in environments ranging from 0°C – 50°C and 0 – 90% non-condensing humidity.
-
Is the Ripen It! Gas Analyzer waterproof?
-
The Store It! Gas Analyzer is not designed to be waterproof and should not be used underwater. The Store It! Gas Analyzer is designed to be water-resistant. Use in light rain is fine, however, there are electronics inside of the Store It! Gas Analyzer that can be damaged from water entering the instrument, so it is not recommended to use the Store It! Gas Analyzer during heavy rainfall.
-
What are the operating conditions of this product?
-
The operating conditions of the Ripen It! Gas Analyzer are 0°C – 50°C and 0 – 90% non-condensing humidity.
-
How many measurements can the included SD card hold?
-
The included 16gb SD card can hold approximately 60 million measurements. If saving data every minute, that equates to approximately 40,000 days of measurement before the SD card is full.
-
What types of tubing materials are recommended to avoid outgas interference?
-
Viton and Teflon have successfully been found not to outgas.
-
Does this product come with a warranty?
-
All devices from Felix Instruments – Applied Food Science come with a 12-month warranty at the point of sale. The warranty covers repairing andreplacing defective parts of the instrument that were manufactured directly by Felix Instruments. The warranty does not cover wear and tear, neglect, misuse, accident, or excessive deterioration. Further, unauthorized repair and alteration by an unauthorized party voids the warranty. More information can be found in the operation manual of your instrument.
Have additional questions?
Click the link below to speak with one of our staff scientists about your specific application.
Specifications
Trigger Mode Measurements C2H4
Sensor Type: | Electrochemical |
Range: | 0-10 ppm |
Resolution: | 0.01 ppm |
Lower Detection Limit: | 0.15 ppm |
Sampling Volume: | 30 ml |
Response Time: | 30 seconds |
Offset Calibration: | Weekly |
Sensor Replaceable (by user): |
yes |
Accuracy (absolute): | 0.15 ppm |
Accuracy (relative): | 5% |
Span Calibration: | 3 months |
Lifetime: | 2 years |
Trigger Mode Measurements CO2
Sensor Type: | Infrared Sensor, Pyroelectric detector |
Range: | 0-100% |
Resolution: | 0.01% |
Lower Detection Limit: | 0.50% |
Sampling Volume: | 12 ml |
Response Time: | 12 seconds |
Offset Calibration: | Weekly |
Sensor Replaceable (by user): |
No |
Accuracy (absolute): | 0.50% |
Accuracy (relative): | 3% |
Span Calibration: | 12 months |
Lifetime: | > 15 years |
Trigger Mode Measurements O2
Sensor Type: | Electrochemical |
Range: | 0-100% |
Resolution: | 0.10% |
Lower Detection Limit: | 0.30% |
Sampling Volume: | 12 ml |
Response Time: | 12 seconds |
Offset Calibration: | Weekly |
Sensor Replaceable (by user): |
Yes |
Accuracy (absolute): | 0.30% |
Accuracy (relative): | 2% |
Span Calibration: | 6 months |
Lifetime: | 2 years |
Continuous Mode Measurements C2H4
Sensor Type: | Electrochemical |
Range: | 0-10 ppm |
Resolution: | 0.010 ppm |
Lower Detection Limit: | 0.030 ppm |
Response Time: | 70 seconds |
Offset Calibration: |
Weekly |
Sensor Replaceable (by user): |
Yes |
Accuracy (absolute): | 0.030 ppm |
Accuracy (relative): | 5% |
Span Calibration: | 3 months |
Lifetime: | 2 years |
Continuous Mode Measurements CO2
Sensor Type: | Infrared Sensor, Pyroelectric detector |
Range: | 0-100% |
Resolution: | 0.01% |
Lower Detection Limit: | 0.01% |
Response Time: | 6 seconds |
Offset Calibration: |
Weekly |
Sensor Replaceable (by user): |
No |
Accuracy (absolute): | 0.01% |
Accuracy (relative): | 3% |
Span Calibration: | 12 months |
Lifetime: | 5 years |
Continuous Mode Measurements O2
Sensor Type: | Electrochemical |
Range: | 0-100% |
Resolution: | 0.10% |
Lower Detection Limit: | 0.10% |
Response Time: | 12 seconds |
Offset Calibration: | Weekly |
Sensor Replaceable (by user): |
Yes |
Accuracy (absolute): | 0.10% |
Accuracy (relative): | 2% |
Span Calibration: | 6 months |
Lifetime: | 2 years |
F-940 Set-Zero (O2)
F-940 Set-Zero (CO2)
F-940 Span Calibration
Standard Gas Tank Attachment Tutorial
Felix Instruments Handheld Gas Analyzer Series Live Training
Increase shelf-life and prevent spoilage.
Rapid, precise measurement of ethylene, CO2 and O2
F-900DK Dynamic Sampling Kit
The Dynamic Sampling Kit allows users to take continuous measurements in closed environments with any of our handheld gas analyzers. The kit includes a 2L acrylic chamber (14cm diameter 12cm deep), wand attachment to connect chamber to the handheld gas analyzer, and battery pack to power fans in the chamber. Additional chamber option: 8L Chamber (24cm diameter 17cm deep).
9XX BE Battery Eliminator for Handheld Gas Analyzers
The Battery Eliminator allows for instruments to operate on AC power. Compatible with F-920, F-940, F-950, and F-960 gas analyzers.