F-940 Store It! Gas Analyzer

Rapid, precise measurement of ethylene, CO2 and O2

Optimize your headspace and MAP analysis workflows.

Gather precise CO2 and O2 data in seconds with the Check It! Gas Analyzer.

F-940 Store It! Gas Analyzer

Backlit LCD screen

Never struggle to see your readings, even in dimly-lit storage environments.

Lightweight & heavy duty

Individually machined in the USA, the Store It! Gas Analyzer’s highly durable, aluminum construction gives you peace of mind without sacrificing weight.

Automated data capture

Never lose a data point with up-to-the-second autosave.

Internal data logging

Store and review thousands of records on the Check It! Gas Analyzer’s removable 16gb SD card.

F-940 Store It! Gas Analyzer
1

Backlit LCD screen

Never struggle to see your readings, even in dimly-lit storage environments.

2

Lightweight & heavy duty

Individually machined in the USA, the Store It! Gas Analyzer’s highly durable, aluminum construction gives you peace of mind without sacrificing weight.

3

Automated data capture

Never lose a data point with up-to-the-second autosave.

4

Internal data logging

Store and review thousands of records on the Check It! Gas Analyzer’s removable 16gb SD card.

Request a live demonstration

Vital QA data collection in the palm of your hand

All-day battery life

With an 8.5 hour battery life and rechargeable 3400 mAh replacement batteries included, you can work confidently without interruption.

Painless data transfer

With bluetooth and WIFI capabilities, rapid data transfer has never been easier.

Rapid measurement

With CO2 and O2 measurements in under 10 seconds, you won’t be stuck waiting for your data.

User-selected sampling modes

Choose from headspace or continuous sampling depending on your unique needs.

5 data points recorded

Each sampling event records CO2 concentration, O2 concentration, date & time, RH, and GPS location.

Wide temperature and humidity range

Gather reliable data in environments ranging from 0°C – 50°C and 0 – 90% non-condensing humidity.

  • Does this product come with a warranty?

  • All devices from Felix Instruments – Applied Food Science come with a 12-month warranty at the point of sale. The warranty covers repairing andreplacing defective parts of the instrument that were manufactured directly by Felix Instruments. The warranty does not cover wear and tear, neglect, misuse, accident, or excessive deterioration. Further, unauthorized repair and alteration by an unauthorized party voids the warranty. More information can be found in the operation manual of your instrument.

  • Is the Ripen It! Gas Analyzer waterproof?

  • The Store It! Gas Analyzer is not designed to be waterproof and should not be used underwater. The Store It! Gas Analyzer is designed to be water-resistant. Use in light rain is fine, however, there are electronics inside of the Store It! Gas Analyzer that can be damaged from water entering the instrument, so it is not recommended to use the Store It! Gas Analyzer during heavy rainfall.

  • What are the operating conditions of this product?

  • The operating conditions of the Ripen It! Gas Analyzer are 0°C – 50°C and 0 – 90% non-condensing humidity.

  • How many measurements can the included SD card hold?

  • The included 16gb SD card can hold approximately 60 million measurements. If saving data every minute, that equates to approximately 40,000 days of measurement before the SD card is full.

  • What types of tubing materials are recommended to avoid outgas interference?

  • Viton and Teflon have successfully been found not to outgas.

Have additional questions?

Click the link below to speak with one of our staff scientists about your specific application.

Specifications

Trigger Mode Measurements C2H4

Sensor Type: Electrochemical
Range: 0-10 ppm
Resolution: 0.01 ppm
Lower Detection Limit: 0.15 ppm
Sampling Volume: 30 ml
Response Time: 30 seconds

Offset Calibration: WeeklySensor Replaceable (by user): yes

Accuracy (absolute): 0.15 ppm
Accuracy (relative): 5%
Span Calibration: 3 months
Lifetime: 2 years

Trigger Mode Measurements CO2

Sensor Type: Infrared Sensor, Pyroelectric detector
Range: 0-100%
Resolution: 0.01%
Lower Detection Limit: 0.50%
Sampling Volume: 12 ml
Response Time: 12 seconds

Offset Calibration: WeeklySensor Replaceable (by user): no

Accuracy (absolute): 0.50%
Accuracy (relative): 3%
Span Calibration: 12 months
Lifetime: 5 years

Trigger Mode Measurements O2

Sensor Type: Electrochemical
Range: 0-100%
Resolution: 0.10%
Lower Detection Limit: 0.30%
Sampling Volume: 12 ml
Response Time: 12 seconds

Offset Calibration: WeeklySensor Replaceable (by user): yes

Accuracy (absolute): 0.30%
Accuracy (relative): 2%
Span Calibration: 6 months
Lifetime: 2 years

Continuous Mode Measurements C2H4

Sensor Type: Electrochemical
Range: 0-10 ppm
Resolution: 0.010 ppm
Lower Detection Limit: 0.030 ppm
Response Time: 70 seconds

Offset Calibration: WeeklySensor Replaceable (by user): yes

Accuracy (absolute): 0.030 ppm
Accuracy (relative): 5%
Span Calibration: 3 months
Lifetime: 2 years

Continuous Mode Measurements CO2

Sensor Type: Infrared Sensor, Pyroelectric detector
Range: 0-100%
Resolution: 0.01%
Lower Detection Limit: 0.01%
Response Time: 6 seconds

Offset Calibration: WeeklySensor Replaceable (by user): no

Accuracy (absolute): 0.01%
Accuracy (relative): 3%
Span Calibration: 12 months
Lifetime: 5 years

Continuous Mode Measurements O2

Sensor Type: Electrochemical
Range: 0-100%
Resolution: 0.10%
Lower Detection Limit: 0.10%
Response Time: 12 seconds

Offset Calibration: WeeklySensor Replaceable (by user): yes

Accuracy (absolute): 0.10%
Accuracy (relative): 2%
Span Calibration: 6 months
Lifetime: 2 years

F-940 Set-Zero (O2)

The F-940 Store It! Gas Analyzer provides rapid and precise measurements of ethylene, CO2 and O2.

F-940 Set-Zero (CO2)

The Store It! was designed for Quality Assurance staff to increase shelf-life and prevent spoilage by monitoring ripening gases throughout the storage process.

F-940 Span Calibration

The Store It! features an intuitive user interface, making it simple to operate right out of the box.

Standard Gas Tank Attachment Tutorial

A built-in data logger stores thousands of data points and conveniently communicates data through USB or Bluetooth onto a PC or tablet for later analysis.

Increase shelf-life and prevent spoilage.

Rapid, precise measurement of ethylene, CO2 and O2