Resources & Support

Select a product below to view support articles, videos, manuals, and obtain the latest updates.

Operations Manual
DOC Document (.doc)

How to translate a manual into another language

To translate a manual into another language, follow the steps below:

  • Once the manual is open, select the contextual menu by selecting the “…” option in the upper-right hand corner of your screen.
  • Select the translate option.
  • Select your preferred language. Select translate.

Note: Due to limitations of Microsoft Translate, manuals greater than 50 pages in length currently do not support dynamic translation.

Specifications

F-920 Check It! Gas Analyzer

Sampling Flow Rate: 70 mL/min
Total Sampling Time: 10 seconds
Data Saving: Automated, 1 second intervals
Display: Backlit visible transflective LCD
Operating Environment: 0° C – 50° C (0-90 % humidity non-condensing)
Operating Atmospheric Pressure: 500-1200 mbar
Dimensions: 18 cm x 13.5 cm x 5.5 cm
Weight: 0.95 kg
Enclosure: Powder coated aluminum
Battery Life: 8.5 hours
Power Source: Removable rechargeable lithium-ion battery
PC Interface: USB and WI-FI SD card + Bluetooth
Data Recorded with Each Measurement: CO2 and O2 concentrations, date, time, RH, GPS location

Trigger Mode Measurements CO2

Sensor Type: Infrared Sensor, Pyroelectric detector
Range: 0-100%
Resolution: 0.01%
Lower Detection Limit: 0.50%
Sampling Volume: 10 ml
Response Time: 10 seconds
Offset Calibration: Weekly
Sensor Replaceable (by user):
No
Accuracy (absolute): 0.50%
Accuracy (relative): 3%
Span Calibration: 12 months
Lifetime: > 15 years

Trigger Mode Measurements O2

Sensor Type: Electrochemical
Range: 0-100%
Resolution: 0.10%
Lower Detection Limit: 0.30%
Sampling Volume: 10 ml
Response Time: 10 seconds
Offset Calibration:
Weekly
Sensor Replaceable (by user):
Yes
Accuracy (absolute): 0.30%
Accuracy (relative): 2%
Span Calibration: 6 months
Lifetime: 2 years

Continuous Mode Measurements CO2

Sensor Type: Infrared Sensor, Pyroelectric detector
Range: 0-100%
Resolution: 0.01%
Lower Detection Limit: 0.01%
Response Time: 10 seconds
Offset Calibration: Weekly
Sensor Replaceable (by user):
No
Accuracy (absolute): 0.01%
Accuracy (relative): 3%
Span Calibration: 12 months
Lifetime: > 15 years

Continuous Mode Measurements O2

Sensor Type: Electrochemical
Range: 0-100%
Resolution: 0.10%
Lower Detection Limit: 0.10%
Response Time: 5 seconds
Offset Calibration: Weekly
Sensor Replaceable (by user):
Yes
Accuracy (absolute): 0.10%
Accuracy (relative): 2%
Span Calibration: 6 months
Lifetime: 2 years

Felix Instruments Handheld Gas Analyzer Series Live Training

F-920 Firmware & Software Downloads

Version Download Installation Instructions Release Notes
1.9.7.9 FW_9x_v1.9.7.9.dfu release_notes_f-920.html
1.9.7.6 (May 2020) FW_9x_v1.9.7.6.dfu Compatible with all F-920, F-940, F-950, F-960. release_notes_f-920.html
1.9.7.4 (Nov 2019) FW_9x_v1.9.7.4.dfu release_notes_f-920.html
1.9.6.7 (Apr 2019) FW_9x_v1.9.6.7.dfu release_notes_f-920.html
1.9.6.5 (Feb 2019) FW_9x_v1.9.6.5.dfu release_notes_f-920.html
1.9.5.8 (Dec 2017) FW_9x_v1.9.5.8.dfu release_notes_f-920.html
1.9.5.4 (Aug 2017) FW_9x_v1.9.5.4.dfu release_notes_f-920.html

 

G.A.S Software

Software used for real-time monitoring and control of the F-920, as well as recalibration of F-920 units.

G.A.S Software F-9x_G_A_S_v0.61Beta.msi

 

  • Does this product come with a warranty?

  • All devices from Felix Instruments – Applied Food Science come with a 12-month warranty at the point of sale. The warranty covers repairing and replacing defective parts of the instrument that were manufactured directly by Felix Instruments. The warranty does not cover wear and tear, neglect, misuse, accident, or excessive deterioration. Further, unauthorized repair and alteration by an unauthorized party voids the warranty. More information can be found in the operation manual of your instrument.

  • Is the Produce Quality Meter waterproof?

  • The Check It! Gas Analyzer is not designed to be waterproof and should not be used underwater. The Check It! Gas Analyzer is designed to be water-resistant. Use in light rain is fine, however, there are electronics inside of the Check It! Gas Analyzer that can be damaged from water entering the instrument, so it is not recommended to use the Check It! Gas Analyzer during heavy rainfall.

  • What are the operating conditions of this product?

  • The operating conditions of the Produce Quality Meter are 0-45° Celsius and 0% – 90% noncondensing humidity.

  • How many measurements can the included SD card hold?

  • The included 16gb SD card can hold approximately 60 million measurements. If saving data every minute, that equates to approximately 40,000 days of measurement before the SD card is full.

  • What types of tubing materials are recommended to avoid outgas interference?

  • Viton and Teflon have successfully been found not to outgas.

Loading the form...

Still need help?