Resources & Support
Select a product below to view support articles, videos, manuals, and obtain the latest updates.
Specifications
F-901R AccuRipe
Sampling Flow Rate: | 70 mL/min |
Total Sampling Time: | 80 seconds |
Data Saving: | Automated, 1 second intervals |
Display: | Sunlight visible transflective LCD |
Operating Environment: | 0° C – 50° C (0-90 % humidity non-condensing) |
Operating Atmospheric Pressure: | 500-1200 mbar |
Dimensions: | 18 cm x 13.5 cm x 5.5 cm |
Weight: | 0.95 kg |
Enclosure: | Powder coated aluminum |
Battery Life: | 8.5 hours |
Power Source: | Removable rechargeable lithium-ion battery |
PC Interface: | USB and WI-FI SD card + Bluetooth |
Data Recorded with Each Measurement: | Ethylene, CO2 and O2 concentrations, date, time, RH, GPS location |
Trigger Mode Measurements C2H4
Sensor Type: | Electrochemical |
Range: | 0-200 ppm |
Resolution: | 0.1 ppm |
Lower Detection Limit: | 0.15 ppm |
Sampling Volume: | 35 ml |
Response Time: | 35 seconds |
Accuracy (absolute): | 2 ppm |
Accuracy (relative): | 5% | Offset Calibration: | Weekly |
Span Calibration: | 3 months | Sensor Replaceable (by user): | yes |
Lifetime: | 2 years |
Trigger Mode Measurements CO2
Sensor Type: | Infrared Sensor, Pyroelectric detector |
Range: | 0-100% |
Resolution: | 0.01% |
Lower Detection Limit: | 0.50% |
Sampling Volume: | 12 ml |
Response Time: | 12 seconds |
Accuracy (absolute): | 0.50% |
Accuracy (relative): | 3% | Offset Calibration: | Weekly |
Span Calibration: | 12 months | Sensor Replaceable (by user): | no |
Lifetime: | 5 years |
Trigger Mode Measurements O2
Sensor Type: | Electrochemical |
Range: | 0-100% |
Resolution: | 0.10% |
Lower Detection Limit: | 0.30% |
Sampling Volume: | 12 ml |
Response Time: | 12 seconds |
Accuracy (absolute): | 0.30% |
Accuracy (relative): | 2% | Offset Calibration: | Weekly |
Span Calibration: | 6 months | Sensor Replaceable (by user): | yes |
Lifetime: | 2 years |
Continuous Mode Measurements C2H4
Sensor Type: | Electrochemical |
Range: | 0-200 ppm |
Resolution: | 0.1 ppm |
Lower Detection Limit: | 0.2 ppm |
Response Time: | 80 seconds |
Accuracy (absolute): | 0.2 ppm |
Accuracy (relative): | 5% | Offset Calibration: | Weekly |
Span Calibration: | 3 months | Sensor Replaceable (by user): | yes |
Lifetime: | 2 years |
Continuous Mode Measurements CO2
Sensor Type: | Infrared Sensor, Pyroelectric detector |
Range: | 0-100% |
Resolution: | 0.01% |
Lower Detection Limit: | 0.01% |
Response Time: | 6 seconds |
Accuracy (absolute): | 0.01% |
Accuracy (relative): | 3% | Offset Calibration: | Weekly |
Span Calibration: | 12 months | Sensor Replaceable (by user): | no |
Lifetime: | 5 years |
Continuous Mode Measurements O2
Sensor Type: | Electrochemical |
Range: | 0-100% |
Resolution: | 0.10% |
Lower Detection Limit: | 0.10% |
Response Time: | 12 seconds |
Accuracy (absolute): | 0.10% |
Accuracy (relative): | 2% | Offset Calibration: | Weekly |
Span Calibration: | 6 months | Sensor Replaceable (by user): | yes |
Lifetime: | 2 years |
F-950 Set-Zero (O2)
The F-950 Three Gas Analyzer measures levels of Ethylene, CO2 and O2 to maintain optimal produce quality at every phase of the supply chain.
F-950 Set-Zero (CO2)
The F-950’s versatile design allows it to be be scaled to many environments, from cold storage to transportation container.
F-950 Set-Zero (C2H4)
Simple to operate and weighing less than a kilogram, the F-950 uses an electrochemical cell to measure ethylene between 0-200 ppm.
F-950 Span Calibration
The F-950 is ideal for measuring ethylene production across a wide range of fruit types.
Standard Gas Tank Attachment Tutorial
Versatile ethylene measurement from 0-200 ppm.
F-950 Three Gas Analyzer
Versatile ethylene measurement from 0-200 ppm.
Current Version
Firmware
The latest release of firmware for the F-950.
*Hardware Version | Firmware Version |
---|---|
4 and greater | 1.9.7.6 |
Less than 4 | 1.7.5.8 |
*Hardware version is the middle digit of the Serial Number. For example, if the serial number is F-950-004-17001 then the hardware version is 4. Check the back of your instrument to find your serial number
Software
The “F-950 Package Setup” installs the latest versions of the Firmware Update Tool and F-950 Controller.
Firmware Update Tool
Allows for installation of the newest firmware releases onto F-950 units.
The latest firmware version can be downloaded below.
Controller
*It is recommended to use G.A.S Software instead of Controller (can be downloading by clicking link below)
Software used for real-time monitoring and control of the F-950, as well as recalibration of F-950 units.
G.A.S Software
This software is only compatible for hardware version 4 and higher.
Release Notes
The release notes provide specific information about what changes there are in the newest firmware version.
-
Does this product come with a warranty?
-
All devices from Felix Instruments – Applied Food Science come with a 12-month warranty at the point of sale. The warranty covers repairing and replacing defective parts of the instrument that were manufactured directly by Felix Instruments. The warranty does not cover wear and tear, neglect, misuse, accident, or excessive deterioration. Further, unauthorized repair and alteration by an unauthorized party voids the warranty. More information can be found in the operation manual of your instrument.
-
Is the Three Gas Analyzer waterproof?
-
The Three Gas Analyzer is not designed to be waterproof and should not be used underwater. The Three Gas Analyzer is designed to be water-resistant. Use in light rain is fine, however, there are electronics inside of the Three Gas Analyzer that can be damaged from water entering the instrument, so it is not recommended to use the Three Gas Analyzer during heavy rainfall
-
What are the operating conditions of this product?
-
The operating conditions of the Ripen It! Gas Analyzer are 0°C – 50°C and 0 – 90% non-condensing humidity.
-
How many measurements can the included SD card hold?
-
The included 16gb SD card can hold approximately 60 million measurements. If saving data every minute, that equates to approximately 40,000 days of measurement before the SD card is full.
-
What types of tubing materials are recommended to avoid outgas interference?
-
Viton and Teflon have successfully been found not to outgas.