Specifications

F-901R AccuRipe

Sampling Flow Rate: 70 mL/min
Total Sampling Time: 80 seconds
Data Saving: Automated, 1 second intervals
Display: Sunlight visible transflective LCD
Operating Environment: 0° C – 50° C (0-90 % humidity non-condensing)
Operating Atmospheric Pressure: 500-1200 mbar
Dimensions: 18 cm x 13.5 cm x 5.5 cm
Weight: 0.95 kg
Enclosure: Powder coated aluminum

Battery Life: 8.5 hours
Power Source: Removable rechargeable lithium-ion battery
PC Interface: USB and WI-FI SD card + Bluetooth
Data Recorded with Each Measurement: Ethylene, CO2 and O2 concentrations, date, time, RH, GPS location

Trigger Mode Measurements C2H4

Sensor Type: Electrochemical
Range: 0-200 ppm
Resolution: 0.1 ppm
Lower Detection Limit: 0.15 ppm
Sampling Volume: 35 ml
Response Time: 35 seconds
Accuracy (absolute): 2 ppm
Accuracy (relative): 5%
Offset Calibration: Weekly
Span Calibration: 3 months
Sensor Replaceable (by user): yes
Lifetime: 2 years

Trigger Mode Measurements CO2

Sensor Type: Infrared Sensor, Pyroelectric detector
Range: 0-100%
Resolution: 0.01%
Lower Detection Limit: 0.50%
Sampling Volume: 12 ml
Response Time: 12 seconds
Accuracy (absolute): 0.50%
Accuracy (relative): 3%
Offset Calibration: Weekly
Span Calibration: 12 months
Sensor Replaceable (by user): no
Lifetime: 5 years

Trigger Mode Measurements O2

Sensor Type: Electrochemical
Range: 0-100%
Resolution: 0.10%
Lower Detection Limit: 0.30%
Sampling Volume: 12 ml
Response Time: 12 seconds
Accuracy (absolute): 0.30%
Accuracy (relative): 2%
Offset Calibration: Weekly
Span Calibration: 6 months
Sensor Replaceable (by user): yes
Lifetime: 2 years

Continuous Mode Measurements C2H4

Sensor Type: Electrochemical
Range: 0-200 ppm
Resolution: 0.1 ppm
Lower Detection Limit: 0.2 ppm
Response Time: 80 seconds
Accuracy (absolute): 0.2 ppm
Accuracy (relative): 5%
Offset Calibration: Weekly
Span Calibration: 3 months
Sensor Replaceable (by user): yes
Lifetime: 2 years

Continuous Mode Measurements CO2

Sensor Type: Infrared Sensor, Pyroelectric detector
Range: 0-100%
Resolution: 0.01%
Lower Detection Limit: 0.01%
Response Time: 6 seconds
Accuracy (absolute): 0.01%
Accuracy (relative): 3%
Offset Calibration: Weekly
Span Calibration: 12 months
Sensor Replaceable (by user): no
Lifetime: 5 years

Continuous Mode Measurements O2

Sensor Type: Electrochemical
Range: 0-100%
Resolution: 0.10%
Lower Detection Limit: 0.10%
Response Time: 12 seconds
Accuracy (absolute): 0.10%
Accuracy (relative): 2%
Offset Calibration: Weekly
Span Calibration: 6 months
Sensor Replaceable (by user): yes
Lifetime: 2 years

F-950 Set-Zero (O2)

The F-950 Three Gas Analyzer measures levels of Ethylene, CO2 and O2 to maintain optimal produce quality at every phase of the supply chain.

F-950 Set-Zero (CO2)

The F-950’s versatile design allows it to be be scaled to many environments, from cold storage to transportation container.

F-950 Set-Zero (C2H4)

Simple to operate and weighing less than a kilogram, the F-950 uses an electrochemical cell to measure ethylene between 0-200 ppm.

F-950 Span Calibration

The F-950 is ideal for measuring ethylene production across a wide range of fruit types.

Standard Gas Tank Attachment Tutorial

Versatile ethylene measurement from 0-200 ppm.

F-950 Three Gas Analyzer

Versatile ethylene measurement from 0-200 ppm.

Current Version

Firmware

The latest release of firmware for the F-950.

*Hardware Version
Firmware Version
4 and greater 1.9.7.6
Less than 4 1.7.5.8

*Hardware version is the middle digit of the Serial Number. For example, if the serial number is F-950-004-17001 then the hardware version is 4. Check the back of your instrument to find your serial number

Software

The “F-950 Package Setup” installs the latest versions of the Firmware Update Tool and F-950 Controller.


Firmware Update Tool


Allows for installation of the newest firmware releases onto F-950 units.

The latest firmware version can be downloaded below.

Controller

*It is recommended to use G.A.S Software instead of Controller (can be downloading by clicking link below)

Software used for real-time monitoring and control of the F-950, as well as recalibration of F-950 units.


G.A.S Software


This software is only compatible for hardware version 4 and higher.

Release Notes

The release notes provide specific information about what changes there are in the newest firmware version.

Firmware: v1.9.7.6 (May 2020)

FW_9x_v1.9.7.6.dfu

Firmware: v1.9.7.4 (Nov 2019)

FW_9x_v1.9.7.4.dfu

F-950 Package Setup

F-950_Package_Setup_v1.2.exe

Firmware: 1.7.5.8 (April 2017)

FW_950_v1.7.5.8.dfu

Installation Instructions in PDF form

F_950_Package_Installation_Instructions.pdf

F-9x G.A.S Software

F-9x_G_A_S_v0.61Beta.msi

Firmware Upgrade Instructions

F_950_Firmware_Upgrade_Instructions.pdf

  • Does this product come with a warranty?

  • All devices from Felix Instruments – Applied Food Science come with a 12-month warranty at the point of sale. The warranty covers repairing and replacing defective parts of the instrument that were manufactured directly by Felix Instruments. The warranty does not cover wear and tear, neglect, misuse, accident, or excessive deterioration. Further, unauthorized repair and alteration by an unauthorized party voids the warranty. More information can be found in the operation manual of your instrument.

  • Is the Three Gas Analyzer waterproof?

  • The Three Gas Analyzer is not designed to be waterproof and should not be used underwater. The Three Gas Analyzer is designed to be water-resistant. Use in light rain is fine, however, there are electronics inside of the Three Gas Analyzer that can be damaged from water entering the instrument, so it is not recommended to use the Three Gas Analyzer during heavy rainfall

  • What are the operating conditions of this product?

  • The operating conditions of the Ripen It! Gas Analyzer are 0°C – 50°C and 0 – 90% non-condensing humidity.

  • How many measurements can the included SD card hold?

  • The included 16gb SD card can hold approximately 60 million measurements. If saving data every minute, that equates to approximately 40,000 days of measurement before the SD card is full.

  • What types of tubing materials are recommended to avoid outgas interference?

  • Viton and Teflon have successfully been found not to outgas.


Enter your full name


Enter your email address


Enter your phone number


Fill the name of your company


Fill the country where you need support



Tick if you currently have an instrument


Select the product you own from the drop down list


Enter the serial number of the owned product


Tick if you are looking for software support


Select the software








Still need help?